●120 μm pixel pitch
●Excellent sensitivity and high dynamic range modes
●Suitable for dental CBCT application, ensures the overall imaging of the mouth and jaw
●ROI area scan function can replace the linear array detector
●CsI scintillator offers lower dose and high image quality
X-Panel 1917z FDM is a low noise X-ray flat panel detector optimized for dental CBCT application. It is based on Indium Gallium Zinc Oxide (IGZO) technology that supports a fast frame rate among other benefits. X-Panel 1917z FDM has excellent image quality and large dynamic range. Furthermore, it comes with multiple gain modes, that make it compatible with both high sensitivity and large dynamic range requirements.
Technology |
Sensor |
IGZO |
Scintillator |
CsI |
Active Area |
184 x 166 mm |
Pixel Matrix |
1536 x 1386 |
Pixel Pitch |
120 μm |
AD Conversion |
16 bits |
Interface |
Data Interface |
2.5G Ethernet |
Exposure Control |
Pulse Sync In / Pulse Sync Out |
Frame Speed |
40 fps (1×1) |
Operating System |
Windows7 / Windows10 OS 32 bits or 64 bits |
Technical Performance |
Resolution |
4.0 lp/mm |
Energy Range |
40~160 KV |
Lag |
0.8% @ 1st frame |
Dynamic Range |
≥86 dB |
Sensitivity |
460 lsb/uGy |
SNR |
49 dB @(20000lsb) |
MTF |
76% @(1 lp/mm) |
47% @(2 lp/mm) |
28% @(3 lp/mm) |
DQE |
63% @(0 lp/mm) |
42% @(1 lp/mm) |
29% @(2 lp/mm) |
Mechanical |
Dimension(H x W x D) |
225 x 193.2 x 15 mm |
Weight |
2.0 Kg |
Sensor Protection Material
|
Carbon Fiber |
Housing Material |
Aluminum Alloy |
Environmental |
Temperature Range |
10~35℃(operating); -10~50℃(storage) |
Humidity |
30~70% RH(non-condensing) |
Vibration |
IEC 60068-2-6 Operation: 0.7G, 10-150Hz |
Shock |
IEC 60068-2-27 Operation: 2G, duration 11ms |
Dust and Water Resistant
|
IPX0 |
Power |
Supply |
100~240 VAC |
Frequency |
50/60 Hz |
Consumption |
12.5W |
|