Technology | |
Sensor | A-Si |
Scintillator | GOS / CSI |
Active Area | 160 x 128 mm |
Pixel Matrix | 1024 x 1248 |
Pixel Pitch | 125 μm |
AD Conversion | 16 bits |
Interface | |
Communication Interface | Gigabit Ethernet |
Exposure Control | Pulse Sync In (Edge or Level) / Pulse Sync Out (Edge or Level) |
Mode | Software Mode / HVG Sync Mode / FPD Sync Mode |
Image Acquisition Time | 30fps |
Operating System | Windows7 / Windows10 OS 32 bits or 64 bits |
Technical Performance | |
Resolution | 4.0 lp/mm |
Energy Range | 40-160 KV |
Lag | 0.8% 1st frame |
Dynamic Range | ≥86dB |
Sensitivity | 460 lsb/uGy |
SNR | 49 dB @(20000lsb) |
MTF | 75% @(1 lp/mm) |
46% @(2 lp/mm) | |
27% @(3 lp/mm) | |
DQE | 58% @(0 lp/mm) |
43% @(1 lp/mm) | |
30% @(2 lp/mm) | |
Mechanical | |
Dimension(HxWxD) | 196 x 162 x 37.5 mm |
Weight | 2.8 Kg |
Sensor Protection Material | Carbon Fiber |
Housing Material | High Strength Aluminum Alloy |
Environmental | |
Temperature Range | 10~35°℃(operating); -10~50℃(storage) |
Humidity | 30~70% RH(non-condensing) |
Vibration | IEC/EN 60721-3 class 2M3(10~150 Hz,0.5 g) |
Shock | IEC/EN 60721-3 class 2M3(11 ms,2 g) |
Dust and Water Resistant | IPX0 |
Power | |
Supply | 100~240 VAC |
Frequency | 50/60 Hz |
Consumption | 8W |
Application | |
Medical |
Dental CBCT |
Mechanical Dimension |
|