Technology | |
Sensor | A-Si |
Scintillator | CSI |
Active Area | 239 x 304 mm |
Pixel Matrix | 2816 x 3584 |
Pixel Pitch | 85 μm |
AD Conversion | 16 bits |
Interface | |
Communication Interface | Gigabit Ethernet |
Exposure Control | Trigger In(Edge or Level) / Trigger Out (Edge or Level) |
Image Acquisition Time | ≤1 seconds |
Operating System | Windows7 / Windows10 OS 32 bits or 64 bits |
Technical Performance | |
Resolution | 5.8 lp/mm |
Energy Range | 40-160 KV |
Lag | ≤1% @1st frame |
Dynamic Range | ≥86dB |
Sensitivity | 67 lsb/uGy |
SNR | 45 dB @(20000lsb) |
MTF | 76% @(1 lp/mm) |
47% @(2 lp/mm) | |
27% @(3 lp/mm) | |
DQE | 60% @(0 lp/mm) |
54% @(1 lp/mm) | |
48% @(2 lp/mm) | |
Mechanical | |
Dimension(H x W x D) | 334 x 273.5 x 34 mm |
Weight | 2.1 Kg |
Sensor Protection Material | Carbon Fiber |
Housing Material | High Strength Aluminum Alloy |
Environmental | |
Temperature Range | 10~35°℃(operating);-10~50℃(storage) |
Humidity | 30~70% RH(non-condensing) |
Vibration | IEC/EN 60721-3 class 2M3(10~150 Hz,0.5 g) |
Shock | IEC/EN 60721-3 class 2M3(11 ms,2 g) |
Dust and Water Resistant | IPX0 |
Power | |
Supply | 100~240 VAC |
Frequency | 50/60 Hz |
Consumption | 12W |
Application | |
Medical |
Mammography |
Mechanical Dimension |
|